Description
Ryujin Plus rapid exchange balloon catheter is intended for use during percutaneous transluminal coronary angioplasty (PTCA) for the purpose of improving myocardial blood flow in the localized stenotic lesion of the coronary arteries.
This device is designed for crossing routine lesions without compromising trackability.
Characteristics
- Flexible distal shaft and hydrophilic coating for good trackability
- The low profile enables the kissing balloon technique in a 6-Fr large lumen (< 0.070″ / 1.80 mm) guiding catheter
- Fixed outer and inner tube prevents accordion folds from forming at the balloon section
- A broad selection of diameters (from 1.25 mm to 4.0 mm) and lengths (from 10 mm to 40 mm)
General specifications
Shaft diameter | 2.0 Fr (0.67 mm) Proximal 2.4 Fr (0.80 mm) Distal for 1.25 mm to 2.0 mm 2.5 Fr (0.83 mm) Distal for 2.25 mm to 3.00 mm 2.6 Fr (0.87 mm) Distal for 3.50 mm and 4.00 mm |
---|---|
Usable length | 145 cm |
Nominal pressure | 6 atm (608 kPa) |
Rated burst pressure | 14 atm (1419 kPa) for 1.25 mm to 3.00 mm 12 atm (1216 kPa) for 3.50 mm and 4.00 mm |
Guidewire compatibility | Maximum diameter 0.014″ (0.36 mm) |
Lesion entry profile | 0.43 mm |
Coating | Hydrophilic |
Reviews